Fume hoods » Laminated
Q-Optimal
This is a series of laminated fume cupboards based on a sturdy frame made of steel sections. The series offers a great cost-effect ratio in situations when fire-proofing of the cupboard construction is not necessary.
A wide selection of versions and sizes is available including unconventional orders, adjusted to clients’ specific requirements.
The fume cupboards on offer include versions with worktop at 900 mm height, with lowered worktop or without one, designed for low-ceiling rooms or double-sided and ones with glazed side walls. We also realize requests involving bench-mounted fume cupboards and special constructions.
A large number of finishing touches and configurations, ease of modification and installing extra appliances ensure the fulfilment of all requirements that may be expected from a modern fume cupboard in a modern laboratory.
The slot ventilation system guarantees even airflow throughout all of the working chamber interior and effective capturing of gases, dust and fumes both lighter or heavier than air, making the working conditions for the operator as safe as possible.
The underbench ventilated cabinet, connected to the fume cupboard ventilation system or an independent one (on request), is a comfortable, convenient space for short-term storage of reagents, equipment or tools needed for operation.
Every fume cupboard is typically equipped with a control system complete with air flow sensor, fully compliant with EN-14175 standard. The fume cupboards are also equipped with two 230V sockets as well as a small sink and two cold water outlets which are operated from the front panel.
Worktop:
- Solid ceramics (standard)
- Large-size Buchtal ceramics
- Durcon epoxy resin
- Polypropylene
Working chamber lining:
- Laminate (standard)
- Large-size Buchtal ceramics
- Mixture of phenolic resins - Max Resistance
- Polypropylene
Sash window:
- MDF frame with sliding glass panes (standard)
- Aluminium frame
- Steel frame
- Tempered glass or polycarbonate (for working with hydrofluoric acid)
Airflow sensor:
- Q-Flow (standard)
- Q-Flow Compact
- Q-Flow Touch
- Q-Flow EX
- Schneider FM- 100
- Schneider FM- 500
- Schneider iCM
- Schneider FM-100E
Standard equipment:
- 2 pcs. 230V sockets
- 2 water outlets with taps on the front panel
- Small sink (approx. 280x80mm)
Additional options:
- Additional 230V and 380V sockets
- Additional water outlets
- Flammable and technical gases outlets
- Differential circuit breaker
- Automatic window control (Auto Protect system)
- Manual Protect System (reminding of the necessity to shut the window)
- Automatic controlling of the ventilation depending on sash position
Type | 1200 | 1500 | 1800 |
Width (mm) | 1270 | 1570 | 1870 |
Depth (mm) | 940 | ||
Height (closed/open sash) (mm) | 2450 / 2500 | ||
Worktop height (mm) | 900 | ||
Working chamber width (mm) depth (mm) height (mm) |
1150 720 1350 |
1450 720 1350 |
1750 720 1350 |
Maximum sash opening (mm) | 720 | ||
Recommended airflow with sash open (m3/h) | 600-1000 | 750-1300 | 950-1500 |
Recommended air flow speed at front face (m/s) | 0,3 - 0,5 | ||
Extract-air manifold (mm) | Ø160 | Ø200 | Ø250 |
Type | 1400 | 1700 | 2000 |
Width (mm) | 1390 | 1690 | 1990 |
Depth (mm) | 940 | ||
Height (closed/open sash) (mm) | 2450 / 2450 | ||
Worktop height (mm) | min. 100 | ||
Working chamber width (mm) depth (mm) height (mm) |
1150 660 max. 2150 |
1450 660 max. 2150 |
1750 660 max. 2150 |
Maximum sash opening (mm) | 1500 | ||
Recommended airflow with sash open (m3/h) | 1800-3000 | 2300-3850 | 2800-4650 |
Recommended air flow speed at front face (m/s) | 0,3 - 0,5 | ||
Extract-air manifold (mm) | Ø315 | Ø315 | 2x Ø250 |
Type | 1200 | 1500 | 1800 |
Width (mm) | 1270 | 1570 | 1870 |
Depth (mm) | 940 | ||
Height (closed/open sash) (mm) | 2100/2100 | ||
Worktop height (mm) | 900 | ||
Working chamber width (mm) depth (mm) height (mm) |
1150 660 1000 |
1450 660 1000 |
1750 660 1000 |
Maximum sash opening (mm) | 650 | ||
Recommended airflow with sash open (m3/h) | 600-1000 | 750-1300 | 950-1550 |
Recommended air flow speed at front face (m/s) | 0,3 - 0,5 | ||
Extract-air manifold (mm) | Ø160 | Ø200 | Ø250 |
Type | 1400 | 1700 | 2000 |
Width (mm) | 1390 | 1690 | 1990 |
Depth (mm) | 940 | ||
Height (closed/open sash) (mm) | 2450 / 2500 | ||
Worktop height (mm) | 900 | ||
Working chamber width (mm) depth (mm) height (mm) |
1150 660 1350 |
1450 660 1350 |
1750 660 1350 |
Maximum sash opening (mm) | 720 | ||
Recommended airflow with sash open (m3/h) | 1200-2000 | 1550-2550 | 1850-3100 |
Recommended air flow speed at front face (m/s) | 0,3 - 0,5 | ||
Extract-air manifold (mm) | Ø250 | Ø315 | Ø315 |